undefined
+
  • undefined

X5 Series Visible Spectrophotometer

At present, the main products of Yi spectrum instrument are: handheld Raman spectrometer, portable Raman spectrometer, micro Raman spectrometer series products; visible spectrophotometer, ultraviolet visible spectrophotometer, quasi double beam ultraviolet visible spectrophotometer, double beam ultraviolet visible spectrophotometer series products;

Product accessories:

Inquiry

Details Introduction


  • Characteristic
  • Function
  • Parameter
  • Configuration
  • Video
  • Accessory
    • Commodity name: X5 Series Visible Spectrophotometer

    Product Features
    ● High-resolution C- T structure monochromator, good stability, reproducibility and accuracy.
    ● Advanced ARM microcomputer technology, excellent electronic circuit and signal processing, stable and reliable data, U disk data storage and export, software one-key upgrade, Bluetooth printing and other data output.
    ●Integrated aviation-grade aluminum alloy substrate, machined mirror seat to ensure long-term stability of the optical path.
    ● 7-inch color large-screen graphic liquid crystal display, with navigation menu interface design, clear operation commands, with special functions to prevent users from operating errors, users can operate without checking the user manual; touch screen + key double operation mode, with gloves can also operate. ● Strict manufacturing process and 32 key quality control points, so that each instrument is very durable.
    ● Large capacity memory, can store massive test data.
    ● Tungsten lamp, photocell, precision screw rod and other key devices are selected from well-known and imported brands to ensure reliable quality of the instrument.
    ● Four-slot 10-50mm universal adjustable manual sample holder, convenient for different specifications of the cuvette test needs.
    ● Independently designed tungsten lamp replacement window for easy replacement.

    Product Function


    1. Photometric Measurement: You can select the single-point test wavelength and test method you need in the range of 320-1100nm to determine the absorbance or transmittance of the sample. The concentration of the sample can also be read directly by inputting the standard sample concentration or concentration factor. 2, quantitative measurement: through the known parameter factor curve or automatically establish the standard solution curve, measure the sample solution of unknown concentration.
    3. Qualitative measurement: set a wavelength range, scan intervals, and then measure the absorbance, transmittance, reflectance and energy of solid or liquid samples at intervals. The measured atlas can also be superimposed, scaled, smoothed, filtered, peak-valley detected, saved, printed, etc.
    4. Time measurement: Time measurement is also called dynamic measurement. The sample is scanned according to the time range interval of absorbance or transmittance at the set wavelength point. The absorbance can also be converted into concentration or reaction rate calculation by inputting the concentration factor.
    Enzyme kinetic reaction rate calculation. Can also be measured out of the map overlay, scaling, smoothing, filtering, peak valley detection, derivation and other map processing methods for your choice.
    5, multi-wavelength measurement: multiple wavelength points can be set, and the absorbance or transmittance of the sample solution can be measured by setting n wavelength test wavelengths to test the absorbance and transmittance ratio at a single wavelength.
    6, instrument calibration measurement: built-in special wavelength larger allowable error calibration, transmission larger allowable error measurement, stray light measurement.
    7. Auxiliary functions: cumulative time of tungsten lamp lighting, independent shutdown and startup of tungsten lamp, operation language selection, automatic wavelength calibration, Bluetooth function, etc.
    8. Optional PC anti-control software package functions: quantitative measurement, qualitative measurement, time measurement, multi-wavelength measurement, 3D graphics analysis, map overlay analysis, etc.

    product parameters

    model

    XV-5

    optical system

    single beam; 1200 strips/mm holographic grating

    wavelength range (nm)

    320-1100

    spectral bandwidth (nm)

    2

    wavelength accuracy (nm)

    ±0.5

    wavelength repeatability (nm)

    ≤ 0.1

    transmittance accuracy (%T)

    ±0.3

    transmittance repeatability (%T)

    ≤ 0.1

    noise (%T)

    0% line 0.2;100% line 0.3

    drift (Abs)

    ≤ 0.001

    baseline flatness (Abs)

    ±0.001

    stray light (%T)

    ≤ 0.05 @ 360nm

    photometric range

    0-200% T,-0.301-3A, 0-9999C(0-9999F)

    test mode

    transmission, reflection, absorbance, concentration, energy

    light source

    deuterium lamp + tungsten lamp

    Display

    7-inch Touch Screen Display (Resolution 800*480)

    System Language

    Chinese, English, Russian, Spanish

    Software Upgrade

    Support U Disk One-click Upgrade

    Data Output

    U Disk, Bluetooth, USB Software Upload

    Data Output Format

    CSV

    Operation Mode

    Touch Screen

    sample rack

    4-slot 10-50mm universal adjustable manual sample rack

    host function

    basic + quantitative + qualitative + time + multi-wavelength

    printer

    selection, support Bluetooth printing

    power range

    AC90 ~ 250V/ 50~60Hz

    size (L×W×H)mm

    470 × 330 × 210

    weight (kg)

    11

    product standard configuration


    ● host 1 ● power cord 1
    ● 4 slots 10-50mm universal adjustable manual sample rack 1
    ● 1cm standard glass cuvette 1 set 4
    ● host instruction manual 1
    ● dust cover 1